A new inexpensive current probe for on‐line leakage current measurement of biomedical devices in hospital environment is described. The prototype is designed to detect and measure leakage currents on the ground wire of the device’s power cord so that its integrity can be monitored in real time. Realized with a sensing coil specially matched to a low‐noise op amp, this probe adds only negligible impedance on the monitored ground lines. From this preliminary study about the device’s metrological performances, a sensitivity of 10 nArms for a current range 1–500 μArms has emerged, together with a mean linearity error of 0.03% and a frequency response flat within 1% of gain from 50 to 2000 Hz.

New ac micro-ammeter for leakage current measurement of biomedical equipment / Branca, Francesco Paolo; DEL PRETE, Zaccaria; Marinozzi, Franco. - In: REVIEW OF SCIENTIFIC INSTRUMENTS. - ISSN 0034-6748. - 64:11(1993), pp. 3303-3307. [10.1063/1.1144295]

New ac micro-ammeter for leakage current measurement of biomedical equipment

BRANCA, Francesco Paolo;DEL PRETE, Zaccaria;MARINOZZI, Franco
1993

Abstract

A new inexpensive current probe for on‐line leakage current measurement of biomedical devices in hospital environment is described. The prototype is designed to detect and measure leakage currents on the ground wire of the device’s power cord so that its integrity can be monitored in real time. Realized with a sensing coil specially matched to a low‐noise op amp, this probe adds only negligible impedance on the monitored ground lines. From this preliminary study about the device’s metrological performances, a sensitivity of 10 nArms for a current range 1–500 μArms has emerged, together with a mean linearity error of 0.03% and a frequency response flat within 1% of gain from 50 to 2000 Hz.
1993
microammeter; ground currents; patient safety
01 Pubblicazione su rivista::01a Articolo in rivista
New ac micro-ammeter for leakage current measurement of biomedical equipment / Branca, Francesco Paolo; DEL PRETE, Zaccaria; Marinozzi, Franco. - In: REVIEW OF SCIENTIFIC INSTRUMENTS. - ISSN 0034-6748. - 64:11(1993), pp. 3303-3307. [10.1063/1.1144295]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/244104
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact