Substrate deformation affects the behavior of many cell types, as for example bone, skeletal muscle and endothelial cells. Nowadays, in vitro tests are widely employed to study the mechanotransduction induced by substrate deformation. The aim of in vitro systems is to properly reproduce the mechanical stimuli sensed by the tissue in the cellular microenvironment. An accurate strain measurement and control is therefore necessary to ensure the cell sensing the proper strain for the entire treatment. Different types of in vitro systems are commercially available or custom made designed; however, none of these devices performs a real-time measurement of the induced strains. In this study, we proposed a uniaxial strain device for in vitro cell stimulation with an innovative real-time strain control. The system was designed to induce sinusoidal waveform stimulation in a huge range of amplitude and frequency, to three silicone chambers stretched by a linear actuator. The real-time strain measurement and control algorithm is based on an optical tracking method implemented in LabView 2015, and it is able adapting the input amplitude to the linear motor, if necessary, hanging the stimulation signal for about 120 ms. A validation of the strain values measured during the real-time tracking algorithm was carried out through a comparison with digital image correlation (DIC) technique. We investigated the influence of number of reference points and image size on the algorithm accuracy. Experimental results showed that the tracking algorithm allowed for a real-time measurement of the membrane longitudinal strains with a relative error of 0.3%, on average, in comparison to the strains measured with DIC in post-processing analysis. We showed a high homogeneity of the strain pattern on the entire chamber base for different stimulation conditions. Finally, as proof of concept, we employed the uniaxial strain device to induce substrate deformation on human Osteosarcoma cell line (SaOS-2). Experimental results showed a consistent cells’ change in shape in response to the mechanical strain.

Development and validation of a device for in vitro uniaxial cell substrate deformation with real-time strain control / Apa, Ludovica; Carraro, Serena; Pisu, Simona; Peruzzi, Barbara; Rizzuto, Emanuele; Del Prete, Zaccaria. - In: MEASUREMENT SCIENCE & TECHNOLOGY. - ISSN 0957-0233. - 31:12(2020). [10.1088/1361-6501/aba011]

Development and validation of a device for in vitro uniaxial cell substrate deformation with real-time strain control

Apa, Ludovica
Primo
;
Carraro, Serena;Pisu, Simona;Rizzuto, Emanuele
;
Del Prete, Zaccaria
Ultimo
2020

Abstract

Substrate deformation affects the behavior of many cell types, as for example bone, skeletal muscle and endothelial cells. Nowadays, in vitro tests are widely employed to study the mechanotransduction induced by substrate deformation. The aim of in vitro systems is to properly reproduce the mechanical stimuli sensed by the tissue in the cellular microenvironment. An accurate strain measurement and control is therefore necessary to ensure the cell sensing the proper strain for the entire treatment. Different types of in vitro systems are commercially available or custom made designed; however, none of these devices performs a real-time measurement of the induced strains. In this study, we proposed a uniaxial strain device for in vitro cell stimulation with an innovative real-time strain control. The system was designed to induce sinusoidal waveform stimulation in a huge range of amplitude and frequency, to three silicone chambers stretched by a linear actuator. The real-time strain measurement and control algorithm is based on an optical tracking method implemented in LabView 2015, and it is able adapting the input amplitude to the linear motor, if necessary, hanging the stimulation signal for about 120 ms. A validation of the strain values measured during the real-time tracking algorithm was carried out through a comparison with digital image correlation (DIC) technique. We investigated the influence of number of reference points and image size on the algorithm accuracy. Experimental results showed that the tracking algorithm allowed for a real-time measurement of the membrane longitudinal strains with a relative error of 0.3%, on average, in comparison to the strains measured with DIC in post-processing analysis. We showed a high homogeneity of the strain pattern on the entire chamber base for different stimulation conditions. Finally, as proof of concept, we employed the uniaxial strain device to induce substrate deformation on human Osteosarcoma cell line (SaOS-2). Experimental results showed a consistent cells’ change in shape in response to the mechanical strain.
2020
substrate deformation; cell stretching; mechanical loading; real-time strain control; optical tracking; mechanotransduction; digital image correlation; SaOS-2.
01 Pubblicazione su rivista::01a Articolo in rivista
Development and validation of a device for in vitro uniaxial cell substrate deformation with real-time strain control / Apa, Ludovica; Carraro, Serena; Pisu, Simona; Peruzzi, Barbara; Rizzuto, Emanuele; Del Prete, Zaccaria. - In: MEASUREMENT SCIENCE & TECHNOLOGY. - ISSN 0957-0233. - 31:12(2020). [10.1088/1361-6501/aba011]
File allegati a questo prodotto
File Dimensione Formato  
Apa_Development-and-validation_2020.pdf

accesso aperto

Note: https://doi.org/10.1088/1361-6501/aba011
Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 3.88 MB
Formato Adobe PDF
3.88 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1443405
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 3
social impact