Abstract PURPOSE: The objective of this investigation is to study prosthodontics and internal components resistance to the masticatory stress and considering different force directions by using Finite Element Method analysis (FEM). The structural materials of the components are usually Titanium alloy grade 4 or 5 and thus, guarantee the integration of the fixture in the bone due to the osteointegration phenomena. Even if the long-term dental implant survival rate is easy to be obtained and confirmed by numerous researches, the related clinical success, due to the alteration of the mechanical and prosthodontics components is still controversial. METHODS: By applying engineering systems of investigations like FEM and Von Mises analyses, it has been investigated how dental implant material was held against the masticatory strength during the dynamic masticatory cycles. A three-dimensional system involved fixture, abutment and the connection screws, which were created and analyzed. The elastic features of the materials used in the study were taken from recent literature data. RESULTS: Data revealed a different response for both types of devices, although implant neck and dental abutment showed better results for all conditions of loading while the abutment screw represented aweak point of the system. CONCLUSION: The data of this virtual model showed all the features of different prosthetic retention systems under the masticatory load. Clinicians should find better prosthetic balance in order to better distribute the stress over the component and to guarantee patients' clinical long-term results.

Fem and Von Mises analysis of OSSTEM ® dental implant structural components: evaluation of different direction dynamic loads / Cervino, G1; Romeo, U; Lauritano, F1; Bramanti, E1; Fiorillo, L1; D'Amico, C1; Milone, D3; Laino, L4; Campolongo, F5; Rapisarda, S1; Cicciù, M1.. - In: THE OPEN DENTISTRY JOURNAL. - ISSN 1874-2106. - 12:(2018), pp. 219-229. [10.2174/1874210601812010219]

Fem and Von Mises analysis of OSSTEM ® dental implant structural components: evaluation of different direction dynamic loads

Romeo U
Secondo
;
2018

Abstract

Abstract PURPOSE: The objective of this investigation is to study prosthodontics and internal components resistance to the masticatory stress and considering different force directions by using Finite Element Method analysis (FEM). The structural materials of the components are usually Titanium alloy grade 4 or 5 and thus, guarantee the integration of the fixture in the bone due to the osteointegration phenomena. Even if the long-term dental implant survival rate is easy to be obtained and confirmed by numerous researches, the related clinical success, due to the alteration of the mechanical and prosthodontics components is still controversial. METHODS: By applying engineering systems of investigations like FEM and Von Mises analyses, it has been investigated how dental implant material was held against the masticatory strength during the dynamic masticatory cycles. A three-dimensional system involved fixture, abutment and the connection screws, which were created and analyzed. The elastic features of the materials used in the study were taken from recent literature data. RESULTS: Data revealed a different response for both types of devices, although implant neck and dental abutment showed better results for all conditions of loading while the abutment screw represented aweak point of the system. CONCLUSION: The data of this virtual model showed all the features of different prosthetic retention systems under the masticatory load. Clinicians should find better prosthetic balance in order to better distribute the stress over the component and to guarantee patients' clinical long-term results.
2018
abutment-implant screw; dental prosthesis; finite element analysis; fixture; implant-supported; osteointegration
01 Pubblicazione su rivista::01a Articolo in rivista
Fem and Von Mises analysis of OSSTEM ® dental implant structural components: evaluation of different direction dynamic loads / Cervino, G1; Romeo, U; Lauritano, F1; Bramanti, E1; Fiorillo, L1; D'Amico, C1; Milone, D3; Laino, L4; Campolongo, F5; Rapisarda, S1; Cicciù, M1.. - In: THE OPEN DENTISTRY JOURNAL. - ISSN 1874-2106. - 12:(2018), pp. 219-229. [10.2174/1874210601812010219]
File allegati a questo prodotto
File Dimensione Formato  
Cervino_Fem_2018.pdf

accesso aperto

Note: https://benthamopen.com/ABSTRACT/TODENTJ-12-219
Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 1.47 MB
Formato Adobe PDF
1.47 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1184388
Citazioni
  • ???jsp.display-item.citation.pmc??? 33
  • Scopus 80
  • ???jsp.display-item.citation.isi??? 65
social impact